The OiW-200 Oil-in-Water Analyzer is a compact UV fluorescence sensor and transducer for the purpose of measuring BTEX, PAHs, and oil in water. This sensor features a robust design allowing for it to be used in corrosive and high-pressure environments.
The sensor can be applied to various applications, including wastewater treatment, environmental monitoring, process monitoring, and more. The sensor can be directly suspended in a medium or installed in an AAI flow cell apparatus.
The transducer is a compact device with the purpose of receiving signals from the oil in water sensor. The transducer uses this information to produce and display measurement values. The transducer can also transduce these measurement values into analog current values (4-20 mA) or communicate them via RS232, RS485, or Profibus. The measurement data can be stored to an internal SD memory card on a set schedule. This device also features two relays. One of these may be used with corresponding concentration alarms, and the other relay may be used to switch on a set schedule. The entire transducer is controlled by an easy-to-use touchscreen display.
When light of a particular wavelength hits particles of oil, the particles will absorb some of that light and re-emit light of a different wavelength. This is known as fluorescence. The OiW-200 analyzer measures the concentration of oil in water by monitoring the level of fluorescence.
A transmitter inside the OiW-200 emits light at about 280 nm. The oil particles in the sample absorb some of this energy and then emit light in a range between 300 and 400 nm. This light is detected by the sensor and the intensity is dependent on the concentration of the oil. Each fluorescence spectrum is unique to the specific composition of the oil mixture. The OiW-200 analyzer is suitable for the measurement of refined oils, BTEX, and PAHs in water.
The sensor can be directly suspended in a medium or installed in an AAI flow cell apparatus.
Note: All performance specifications are subject to the assumption that the sample conditioning system and unit installation are approved by Applied Analytics. For any other arrangement, please inquire directly with Sales.
Sensor Overview | |
Light Source | LED < 300 nm Lifetime: 3 years |
Detector | 300 - 400 nm |
Lens Materials | Fused silica Sapphire (optional) |
Measuring Principle | UV fluorescence |
Measurement Interval | Configurable - 1s minimum |
Measuring Range (Adjustable) | Typical: 0-30 ppm 0-100 ppm 0-500 ppm |
Accuracy | +/- 1% FS |
Sensor Material | Stainless steel 1.4404 Titanium (optional) |
Weight | 0.8 kg |
Dimensions | Length: 150 mm Diameter: 38 mm |
Signal Output | 4-20 mA |
Power Consumption (typical) | 0.5 W |
Voltage Requirements | 10 – 32 VDC |
Interface | Modbus RTU |
Maximum Pressure | 6 bar |
Temperature Range | 0 – 55 °C |
Area Classification | General purpose |
Transducer Overview | |
Voltage Feed | 230 VAC (90-260 VAC) Received power max: 15 W |
Display | LCD Touch Panel: 240 X 128 Pixels |
Operational Temperature Range | -10 to 45 °C |
Weight | 1.35 kg |
Mounting | 5 mm mounting hole and track for DIN-rail |
Housing | Polycarbonate, 235 mm X 185 mm X 119 mm Protection code IP65 |
Input | 1 analog input (sensor) 2 pulse inputs selectable to PNP or NPN (also statically usable), switching current 6 A |
Outputs |
2 current outputs, (4-20 mA), active 1 relay, 24 VDC / 0.5 A (resistive load) 1 relay, 230 VAC / 2A (resistive load) or 24 VDC / 6 A (resistive load) |
Communication Interfaces | RS-232, RS-485, Profibus |