The O2A-100 Analyzer is a real-time monitor of oxygen concentration in a stream using the methodology of paramagnetic detection. Engineered to be compact and simple to use.
Request a QuoteThe O2A-100 can be used to monitor oxygen in the following processes:
O2A-100 takes advantage of the unique paramagnetic susceptibility of oxygen. A magnetic field is produced in the analyzer. A suspended assembly is influenced by a magnetic field. The amount that the assembly is influenced by the magnetic field is directly proportional to the oxygen concentration in the gas. A directly proportional electrical current is incorporated to read the oxygen concentration.
| General | |
| Measurement Principle | Paramagnetic sensor |
| Sample Medium | Gas (Process will need to be filtered to less than 3 microns) |
| Analyzer Calibration | Factory calibrated to specific oxygen ranges |
| Reading Verification | Simple verification with air or sample standard |
| Human Machine Interface | Applied Analytics industrial controller with touch screen display |
| Enclosure | 19" rack-mounted analyzer |
| Measuring Parameters | |
| Repeatability | ±1% of full scale |
| Accuracy | ±0.5% of O2 reading |
| Sample Parameters | |
| Sample Temperature | 0°C to 65°C (32°F to 149°F) |
| Sample Pressure | ±1 barg (14.5 psig) |
| Ambient Conditions | |
| Analyzer Environment | Indoor |
| Ambient Temperature | Standard: 0°C to 40°C (32°F to 95°F) |
| Physical Specifications | |
| Dimensions | 7.09" H x 19" W x 9.78" D (180mm H x 483mm W x 248mm D) |
| Weight | 30 lbs. (14 kg) |
| Utility Requirements | |
| Electrical Requirements | 100-240 VAC |
| Outputs/ Communication | |
| Outputs | 1x galvanically isolated 4-20mA analog output 1x fault relay |