Cl2 in TiCl4 off-gas

pdf

Application Summary

Analysis Point Analyte Typ. Range Suitable Analyzer
TiCl4 chlorination process off-gas Cl2 (chlorine) OMA-300 Process Analyzer 0 - 1000 ppm

Introduction

Titanium dioxide (TiO2) is a widely used white pigment found in coatings, plastics, paper, and a variety of other products. One of the primary methods for producing TiO₂ is the chloride process, in which titanium-bearing raw materials are converted to titanium tetrachloride (TiCl4) using chlorine.

The chlorination reaction produces TiCl₄, which is subsequently purified and oxidized to produce high-purity TiO₂. Chlorine generated during the oxidation step is recovered and recycled back into the chlorination process.

Efficient chlorine utilization is critical to the overall economics and operation of the process. Excess chlorine in the chlorinator off-gas represents unutilized process reagent and can place additional demands on downstream recovery and scrubbing equipment.

Continuous measurement of chlorine in the process off-gas provides operators with real-time information about chlorine utilization and process performance.



Measuring Chlorine in TiCl4 Off-Gas

The chlorinator off-gas can contain chlorine along with TiCl4, carbon oxides, nitrogen, and other process components. Depending on the process location and operating conditions, condensable chlorides and particulate matter may also be present.

Traditional photometric analyzers measure absorbance over a limited wavelength range. In a complex process stream, overlapping absorption from multiple compounds can make it difficult to distinguish chlorine from other UV-absorbing components, like TiCl4.

The OMA-300 uses dispersive UV-Vis spectroscopy to acquire the complete absorption spectrum of the sample. The distinctive spectral structure of Cl2 can then be mathematically isolated from the total sample absorbance.

This full-spectrum approach is particularly valuable in TiCl4 production, where chlorine and TiCl4 both exhibit strong UV absorption. The OMA can use the shape of the individual absorption curves to distinguish the contribution of each component and provide a continuous measurement of chlorine concentration.



Process Optimization

Continuous Cl2 measurement can provide operators with an important indication of chlorination efficiency.

If residual chlorine increases, this may indicate that the process is operating with excess chlorine or that process conditions have changed. Conversely, insufficient chlorine can adversely affect chlorination performance and may contribute to incomplete conversion of titanium-bearing feedstock.

By continuously monitoring residual chlorine, plant operators can optimize chlorine addition while maintaining the desired process conditions.

The measurement can also provide an indication of abnormal process conditions and help identify changes before they result in larger process or environmental problems.



OMA-300 Process Analyzer

The OMA-300 is ideally suited for continuous measurement of chlorine in challenging process streams.

The analyzer uses dispersive UV-Vis absorbance spectroscopy to measure the complete absorption spectrum of the sample. Unlike traditional single-wavelength or broadband photometers, the OMA retains the spectral information needed to differentiate overlapping components.

The analyzer can continuously measure up to five chemical components, allowing additional process constituents to be monitored when required.

A fiber-optic design separates the optical measurement cell from the analyzer electronics. This allows the sample flow cell to be constructed with materials selected specifically for the corrosive process conditions while keeping the sample outside the analyzer enclosure.

For chlorine and TiCl₄ applications, Applied Analytics can provide a customized sample conditioning system designed around the process conditions and required measurement point.



System Benefits

  • Continuously measures Cl₂ using full-spectrum UV-Vis absorbance spectroscopy.
  • Distinguishes Cl₂ from overlapping UV-absorbing process components, like TiCl4.
  • Provides real-time information for chlorine utilization and process optimization.
  • Solid-state optical design with no moving parts.
  • Fiber-optically linked sample flow cell keeps corrosive process gas outside the analyzer enclosure.
  • Custom sample conditioning designed for the TiCl₄ process.
  • Can be configured for simultaneous measurement of multiple process components.
  • Proven OMA technology for demanding chlorine and chloride-process applications.


Conclusion

Efficient chlorine utilization is an important part of TiCl4 production. Continuous monitoring of residual Cl2 in process off-gas gives operators direct, real-time information about chlorination performance and can help optimize chlorine consumption while supporting stable plant operation.

The OMA-300 Process Analyzer combines full-spectrum UV-Vis spectroscopy with application-specific sample conditioning to provide a continuous measurement of Cl2 in the demanding environment of TiCl4 production.

By measuring the complete UV absorption spectrum, the OMA-300 can distinguish chlorine from other UV-absorbing compounds in the process stream, providing a more robust analytical solution than conventional photometric techniques.




Further Reading

Subject Location
OMA-300 Process Analyzer Datasheet
Cl2 in TiCl4 Off-Gas Application Note
Advantage of Collateral Data Technical Note