Analytes | hydroquinone monomethyl ether (MeHQ) |
---|---|
Detector | OMA-300 Process Analyzer |
Process Stream | acrylonitrile |
Zeroing Fluid | air |
Typical Measurement Range | 0-100 ppm |
Polymerization inhibitors are chemicals which stabilize reactive monomers and prevent spontaneous polymerization. If too little inhibitor is added to a reactive monomer, polymerization can occur in pipes or vessels and require mechanical removal. On the other hand, if too much inhibtor is added, the final product may have difficulty polymerizing at the intended stage.
MeHQ is usually added as an inhibitor to acyrolinitrile (a precursor to polyacrylonitrile fibers for the textile industry). The concentration of MeHQ in the monomer needs to be regulated to avoid spontaneous polymerization (to polyacrylonitrile) as well as waste of MeHQ.
MeHQ has a distinct UV absorbance curve which allows the OMA system to easily monitor its real-time concentration. The full-spectrum analysis is critical for differentiating MeHQ absorbance from other absorbing impurities in the sample.
The OMA continuously outputs the MeHQ reading to the plant’s main computer, providing new measurements at 1-5 second intervals for extremely tight process control.
Zeroing the system on pure acrylonitrile is highly impractical due to storage difficulty. Since water mixing with acrylonitrile leads to coating and miscibility effects, water is also not a zeroing option. Air has proven to be an effective and practical zeroing fluid when used with a proper sampling system that minimizes bubble formation.
These spectra were taken by the OMA-300 on calibration standard mixtures of MeHQ in Acrylonitrile:
Absorbance Spectrum of MeHQ in Acrylontrile
1st Derivative of Absorbance Spectrum of MeHQ in Acrylontrile
The OMA readings on calibration standards correlate extremely well with lab readings:
The specifications below represent performance/build of the OMA-300 Process Analyzer in a typical MeHQ application.
For technical details about the OMA-300 Process Analyzer, see the data sheet:
DS-001A: OMA-300 Process Analyzer
All performance specifications are subject to the assumption that the sample conditioning system and unit installation are approved by Applied Analytics. For any other arrangement, please inquire directly with Sales.
Performance Specifications | |
---|---|
Accuracy | hydroquinone monomethyl ether 0-100 ppm: ±0.22 ppm |
Note: Subject to modifications. Specified product characteristics and technical data do not serve as guarantee declarations.
Subject | Type |
---|---|
OMA-300 Process Analyzer | Data sheet |
Advantage of Collateral Data | Technical Note |