APG-710
Hydrogen Analyzer



The APG-710 is an analysis module that operates based on the principle of thermal conductivity. It is optimized for the measurement of Hydrogen gas in a wide variety of applications and stream compositions.

  • Compact form factor.
  • Seamless integration with other AAI analyzers.
  • Low noise.
  • Low maintenance.

Request a Quote
Contact Us

Technology: APG-710

The APG-710 can be operated in two ways:

  • Binary gases: Thermal Conductivity Detector (TCD) with Wheatstone Bridge in a continuous method.
  • Multicomponent gases: TCD with Wheatstone Bridge, coupled with a mol-sieve column in an automatic injection method.

Principle of Operation

Gases have a defined thermal conductivity property based on their molecular structure. For the most part, the thermal conductivity of gases is between 0.01 and 0.03 W/mK. Notable exceptions to this are Helium (0.15 W/mK) and Hydrogen (0.18 W/mK).

This significant difference of an order of magnitude between the thermal conductivity of Hydrogen and most other gases can be exploited for measurement purposes by a TCD.

A TCD is comprised of a wheatstone bridge circuit with two equal legs. One leg is exposed to the sample and one leg is utilized as a reference gas. If the gas is a binary gas, then the reference gas that is utilized is the non-Hydrogen gas that makes up the balance of stream composition. If the gas is a multicomponent gas, then reference that is utilized is the same as the carrier gas.

When Hydrogen is present in the sample stream, the difference in thermal conductivity between the sample gas and the reference gas will cause a net imbalance in the wheatstone bridge, which in turn produces a measurable voltage.

The integral of the magnitude of this voltage with respect to time is proportional to the concentration of Hydrogen in the sample stream. This calculation is performed by the AGP-710-H2 controller, which then outputs a 4-20mA signal proportional to the Hydrogen concentration.



Demo Video

Specifications: APG-710

Note: All performance specifications are subject to the assumption that the sample conditioning system and unit installation are approved by Applied Analytics. For any other arrangement, please inquire directly with Sales.

Detector Type Thermal Conductivity Detector (TCD) with wheatstone bridge
Sample Phase Gas
Analyte Hydrogen
Typical Ranges 0-5%, 0-10%, 0-25%, 0-50%, 0-100%
Modes Continuous or Injection Valve w/ Mol-Sieve Column
Communication 4-20mA Signal