• Products
  • MICROSPEC MCP-200 Infrared Analyzer
MCP-200 Infrared Analyzer Straight View
MICROSPEC MCP-200

Infrared Analyzer

The MCP-200 is a real-time concentration monitor for chemical species that absorb IR radiation. Engineered to be compact and modular, the MCP-200 excels as both a rugged standalone and an integrated component.

Request a Quote

Automated infrared process analysis

  • Fast, continuous reading
    For tight process control
  • No recalibration required
    Measurement stabilized by Auto Zero
  • Proprietary optical cell design
    Specialized path length per application

Excels as standalone or integrated unit

  • Modular design
    Easily integrated into any of our systems
  • Rugged and compact
    Cast aluminum enclosure with SS316 flow cell

Process Conditions

  • Analyze liquid or gas
    Using appropriate sample conditioning system
  • Direct analysis of hot/wet sample
    Accepts high temperature/pressure sample

Cost of Ownership

  • No consumables/reagents
    Other than zero gas
  • Proven low-mainentance
    Only requires periodic optics cleaning
  • Solid state instrument
    Outlasts competing models due to lack of moving parts

Compatibility

  • Hazardous area certifications
    Class I Div 1 & 2; ATEX; GOST...
  • No shelter required
    Rugged instrument suitable for outdoor installation
  • Customizable alarms & outputs
    4-20 mA analog outputs; MODBUS; Hart...

Applications

The MicroSpec MCP-200 is used to measure the following chemicals:

  • carbon monoxide | CO
  • carbon dioxide | CO2
  • ethylene | C2H4
  • methane equivalency | CH4
  • water | H2O

Technology

The MCP-200 uses a technology known as Non-Dispersive Infrared (NDIR) Photometry. In a ruggedized package, the MCP-200 pairs an NDIR photometer (2-10 micron) using a tungsten black body source with a proprietary-design optical flow cell. The path length of the flow cell is specialized per measurement application.

The MCP-200 uses serial communication between the detector and the HMI.

Specifications

Note: All performance specifications are subject to the assumption that the sample conditioning system and unit installation are approved by Applied Analytics. For any other arrangement, please inquire directly with Sales.

Measurement Principle Non-dispersive IR photometry (2 to 10 micron)
Light Source Tungsten black body source
Path Length Application-dependent
Sample Introduction Continuous flow from positive pressure stream
Sample Conditioning Custom design per application
Analyzer Calibration If possible, analyzer is factory calibrated with certified calibration fluids; no re-calibration required after initial calibration; measurement normalized by Auto Zero
Reading Verification Simple verification with samples or neutral density filters
Human Machine Interface Industrial controller with touch-screen LCD display running ECLIPSE™ Software
Data Storage Solid State Drive
OPERATING CONDITIONS
Analyzer Environment Indoor/Outdoor (no shelter required)
Ambient Temperature Standard: 0 to 35 °C (32 to 95 °F)
Optional: -20 to 55 °C (-4 to 131 °F)
To avoid radiational heating, use of a sunshade is recommended for systems installed in direct sunlight.
Sample Temperature Standard: -20 to 50 °C (-4 to 122 °F)
Optional: up to 1000 °C (1832 °F)
Sample Pressure 500 psi
UTILITIES
Electrical 5 VDC, 24 VDC controller
Power Consumption 7 watts (maximum)
OUTPUTS
Standard Outputs 1x galvanically isolated 4-20mA analog output per measured analyte(up to 3; additional available by upgrade)
2x digital outputs for fault and SCS control
Optional Outputs Modbus TCP/IP; RS-232; RS-485; Fieldbus; Profibus; HART;
PHYSICAL SPECIFICATIONS
Dimensions 3.75"H x (path length)W x 3.75"D
Path Length 7" to 17" depending on application
Weight 10-30 lbs depending on path length
Process Connections 1/4” tube fittings
Wetted Materials stainless steel 316L, K7 glass, Viton
PERFORMANCE
Accuracy  ±2% full scale
Repeatability  ±2% full scale
Stability  ±2% full scale / 24 hours

The following certifications are available.

NEC Class I, Division 1, Groups C & D; AEx d IIB T3
CEC Class I, Division 1, Groups C & D; AEx d IIB T3
ATEX ATEX Exp II 2(2) GD
GOST GOST Russian Gosstandart Pattern Approval